Velopharyngeal insufficiency is a condition that occurs when the soft palate fails to create an effective seal with the pharynx during speech and swallowing that requires sufficient intraoral pressure. The underlying cause of this velopharyngeal insufficiency is due to congenital or acquired structural defects in the soft palate. This condition results in inadequate velopharyngeal closure, leading to difficulties with speech articulation and swallowing,1,2 and it can significantly impact the quality of life of patients.3,4
A velopharyngeal obturator prosthesis is used to rehabilitate patients with velopharyngeal insufficiency.5,6 This prosthesis comprises three components: the front section, which resembles a removable dental retainer; the middle section, which connects the dental retainer to the bulb; and the bulb section itself, which extends posteriorly towards the pharynx. The function of the bulb extension is to replace the defective portion and make close contact with the pharyngeal tissues for the proper palatopharyngeal seal during speech and swallowing.7,8
In normal individuals, the soft palate elevates and moves backward during swallowing, while the posterior and lateral pharyngeal walls move forward and inward, respectively, to ensure adequate velopharyngeal closure.9 Simultaneously, the tongue also retracts posteriorly, and the base of the tongue generates sufficient swallowing pressure in the soft palate.10 However, in an individual who has a palatal defect and is rehabilitated with an obturator prosthesis, the swallowing pressure produced by the base of the tongue acts on the bulb portion of the obturator. Consequently, this pressure can induce displacement of the obturator prosthesis, and the performance of the swallowing function might be affected if the obturator is not well retained in its position.
Hence, to perform swallowing function properly, the retention and stability of the obturator prosthesis are important, and the retainer design based on the location of the occlusal rests should be considered carefully.11,12 Moreover, according to one case report presented by Collin et al., dislodgement of the obturator prosthesis could potentially lead to an emergency situation if it obstructs the airway.13 Furthermore, the area where the retainer section and the bulb portion connect has a high potential to fracture, and fracturing of the bulb portion can also lead to airway obstruction.
Therefore, it is crucial to evaluate the effect of swallowing tongue pressure on the biomechanics of a velopharyngeal obturator prosthesis and to determine the most suitable occlusal rest position for this obturator prosthesis. To the best of our knowledge, there are no studies about the biomechanical analysis of velopharyngeal obturator prostheses in patients with soft palate defects.
The purpose of the study is to analyze how swallowing tongue pressure affects the biomechanics of a velopharyngeal obturator prosthesis and compare its displacement across different occlusal rest positions.
The cone-beam computed tomography (CBCT) and magnetic resonance imaging (MRI) data of a 36-year-old female participant were used for this finite element analysis (FEA) study. A three-dimensional (3D) geometric model consisting of the maxilla and teeth was developed based on the CBCT data using modelling software (Mimics 19.0; Materialise NV, Leuven, Belgium). The soft palate and a portion of the pharynx, including the nasopharynx and oropharynx, were also simulated in the model using MRI data (Fig. 1). A simulated alveolar bone crest was created 1 mm below the cementoenamel junction of the teeth. The periodontal ligament was simulated with a thickness of 200 µm,14 and the thickness of the mucosa was estimated to be 3 mm.
The right and left first premolars, as well as the second molars, were prepared for the fabrication of surveyed crowns following 3D printing using the 3D printer (3D printer MAX, Asiga, Sydney, Australia). After that, four prepared abutment teeth were scanned for the fabrication of their respective surveyed crowns using the dental lab scanner (E 4 scanner, 3 shape, Copenhagen, Denmark). Next, two types of surveyed crowns were virtually constructed based on different occlusal rest seat positions: one with mesial occlusal rest seats and the other with distal occlusal rest seats. All rest seats were prepared to the same dimensions, specifically 3 mm in width, 3 mm in length, and 1 mm in depth. Lingual ledges were also inserted in all surveyed crowns.
To simulate a congenital palatal defect, a portion of soft palate was resected (Fig. 1A). Two velopharyngeal obturators were virtually fabricated using Aker’s clasps for all abutment teeth, with different occlusal rest positions for each. The posterior extended portion of the obturator prosthesis was positioned at the level of the hard palate to simulate its effective position during function.15 Two experimental models were subsequently generated based on two different velopharyngeal obturator designs: one with “mesial occlusal rests” (Model 1) (Fig. 2A), and the other with “distal occlusal rests” (Model 2) (Fig. 2B).
A meshing software (Visual-Mesh 18.0, ESI Group, Paris, France) was used to create tetrahedral meshes of all components in each experimental model. Model 1 generated a finite element mesh consisting of 632,847 nodes and 3,444,786 elements. Model 2 produced a mesh with 626,417 nodes and 3,433,595 elements. The material properties for this FEA study were assumed to be homogenous, isotropic, and linearly elastic. Table 1 provides a summary of the material properties of all the components used in this study.16-28
Elastic modulus and Poisson’s ratio for anatomical structures and materials
Materials | Elastic modulus (GPa) | Poisson’s ratio ( |
---|---|---|
Cortical bone | 13.7 | 0.30 |
Enamel | 41 | 0.30 |
Dentin | 18.6 | 0.31 |
PDL | 0.05 | 0.49 |
Palatal mucosa | 0.68 | 0.45 |
Soft palate | 6E-6 | 0.45 |
Pharyngeal tissue | 7E-6 | 0.45 |
Zirconia | 210 | 0.27 |
Cement layer | 8 | 0.33 |
Co-Cr alloy | 211 | 0.30 |
Acrylic resin | 2.7 | 0.35 |
GPa, gigapascal; PDL, periodontal ligament; Co-Cr, cobalt-chromium.
The friction coefficients (µ) of 0.01 and 0.10 were used for the contact behavior between the prosthesis and mucosa and between the occlusal rest and surveyed crown, respectively.29 All nodes on the cutting surface of the maxillary bone were constrained in all directions. To simulate swallowing pressure produced by the base of the tongue, 25 kPa was applied at the surface of the bulb of the obturator prosthesis opposite the base of the tongue.30 The analysis was processed using the Visual-Crash software (Visual-Crash for PAM 18.0, ESI Group, Paris, France), and the solver program (Launch VPS Solver 2020.0, ESI Group) was used to perform the analysis. The maximum von-Mises stress and displacement values of the velopharyngeal obturator prosthesis were analyzed and compared for each experimental model.
Table 2 showed the maximum von-Mises stress and displacement values of the velopharyngeal obturator prosthesis and the abutment teeth for each model.
Maximum von-Mises stress and displacement values for obturator prosthesis and abutment teeth
Model | Maximum von-Mises stress (MPa) | Maximum displacement (μm) | ||
---|---|---|---|---|
Metal frame | Acrylic resin | Prosthesis | Abutment teeth | |
Model 1 (mesial occlusal rests) | 64.9 | 4.3 | 31.3 | 0.42 |
Model 2 (distal occlusal rests) | 54.2 | 4.3 | 33.6 | 0.31 |
MPa, megapascal; μm, micrometre.
Concerning biomechanical stress distribution on the obturator prosthesis, the maximum von-Mises stress in the metal framework was slightly higher in model 1 (64.9 MPa) compared to model 2 (54.2 MPa), with both models having the same maximum stress concentration area at the posterior palatal strap adjacent to the junction of the metal framework and obturator bulb (Fig. 3). However, the acrylic resin obturator bulb showed the same maximum von-Mises stress distribution for both models, with a value of 4.3 MPa.
With regard to the displacement distribution of the obturator prosthesis and abutment teeth, model 1 and model 2 exhibited similar displacement values, with 31.3 µm and 33.6 µm for the prosthesis, and 0.42 µm and 0.31 µm for the abutment teeth, respectively. Fig. 4 illustrates the displacement distribution patterns of the velopharyngeal obturator prosthesis for both models.
The dislodgement of the velopharyngeal obturator prosthesis can be potentially life-threatening for patients with soft palate defects who have been rehabilitated with this type of prosthesis. To ensure its efficient and safe function during speech and swallowing, it is crucial for the velopharyngeal obturator to remain stable in its position. This study investigated the effect of swallowing tongue pressure on the biomechanics of the velopharyngeal obturator prosthesis and determined the appropriate occlusal rest position for this type of prosthesis by comparing displacement values between the mesial and distal occlusal rest positions.
During swallowing mechanism, the pressure exerted can vary depending on the size and viscosity of the bolus, types of swallowing, and anatomical locations in the pharynx.30,31 In this study, a pressure of 25 kPa was used to simulate the pressure produced by the base of the tongue in the oropharynx during swallowing, based on data from the study conducted by Qazi et al.30
Previous literature indicates that most studies on occlusal rest positions have primarily focused on distal extension removable partial dentures. Considering the similarity of the posterior extension of the obturator bulb in the velopharyngeal obturator prosthesis to the distal extension partial denture scenario, it is reasonable to compare the findings of this study to those of previous investigations.
The present study found that the von-Mises stress in the metal framework was slightly higher for the mesial occlusal rests compared to the distal occlusal rests. Similarly, Shahmiri et al. suggested that distal occlusal rests should be utilized in distal extension partial dentures to improve stress distribution in the metal framework and acrylic resin denture base structures.29 In contrast, Zarrati et al. proposed using both mesial and distal occlusal rests in distal extension removable partial dentures.32
Basically, moving the occlusal rests from the mesial to the distal side of the abutment teeth could increase the rotation of the denture base. However, in this study, both mesial and distal occlusal rests showed almost the same prosthesis displacement values. The reason for this is the placement of the indirect retainer, which was identical in both models and had the same distance from the fulcrum of the obturator prosthesis.
In this FEA study, the simulation analysis of the biomechanics of the velopharyngeal obturator prosthesis was restricted to the fully dentulous scenario and only focused on the occlusal rest position. Further research is needed to investigate the biomechanical analysis of the velopharyngeal obturator prosthesis in a fully edentulous condition or with different retainer designs in partially edentulous conditions.
Within the limitations of this study, it can be concluded that swallowing tongue pressure had a minor impact on the biomechanics of the velopharyngeal obturator prosthesis and that distal occlusal rests showed a slightly better biomechanical response compared to mesial occlusal rests. The results of this study could probably assist in retainer design considerations during prosthetic rehabilitation for patients with velopharyngeal insufficiency.
This study was supported by grant no 04-2018-0097 from the SNUDH Research Fund.